Limits of absolute vector magnetometry with nitrogen-vacancy centers in diamond
Dennis Lönard, Isabel Cardoso Barbosa, Stefan Johansson, Jonas Gutsche, Artur Widera
Fig. 6, panel (a)
Voigt fit parameters of the example ODMR resonance of Fig. 5 against MW power (dBm): contrast $C_\text{V}$ (%) on the left axis and FWHM linewidth $\alpha_\text{V}$ (MHz) on the right axis, one marker per power step.
Files
- Data: data.csv
Provenance
This panel was regenerated by running the authors' plotting code and data from their public data deposit (Zenodo: https://doi.org/10.5281/zenodo.20052401, CC BY 4.0). The plot shown here is re-rendered for interactive viewing and analysis. See the paper for the original figure and full scientific context.
License: CC BY 4.0.
Phys. Rev. Res. 8, 013054 (2026) · DOI: 10.1103/28t7-7gcn · arXiv: 2504.20750
Other panels of Fig. 6: (b) (c)